+ Part IV · The body

The body's side.

Where blood or tissue fluid wets a device, the body never meets the polymer itself. A coat of the body's own proteins settles on the material at once, and from then on cells meet that coat and whatever molecules leave the material. Part IV follows the body's responses from the surface outward: the protein coat itself (Chapter 8), the cells that gather around an implant or a particle (Chapter 9), the questions that blood alone raises (Chapter 10), and allergy, irritation and harm far from the point of contact (Chapter 11). Each response is explained first as a mechanism and then through the test that reads it. With Part II's account of what leaves a material, this part answers the first half of the central question, why a plastic that is safe in a water bottle can harm a patient, and the toxicology of Part V rests on it.

08 — The first second

The first second.

+ The questionWhat does the body put on a surface before any cell arrives?

Water first, then proteins

Within milliseconds of blood touching a material, proteins from the blood begin to settle on its surface, Kudzin and colleagues write in a 2026 review of how blood reacts with fibrous polymer materials. On the wall of a tube that carries blood, those proteins are in place before any cell arrives, so every cell that comes later meets them and not the material. Chapter 4 followed molecules out of a material, and this chapter follows the body's own molecules onto it.

Water molecules are among the smallest molecules in a body fluid and by far the most numerous, so they reach an implant's surface before anything larger can. A layer of protein follows within the first few minutes. Barberi and Spriano set out that order in a 2021 review of how proteins attach to titanium. Proteins settle by adsorption, in which molecules are held on a surface, unlike absorption, in which they enter the bulk of a material. The adsorbed proteins together make up the surface's protein coat.

A protein is a long chain of amino acids folded into a particular shape, with most of its water-loving groups on the outside and many of its water-repelling groups tucked inside. It is far larger than a water molecule and reaches the surface later, carried by the same random motion that Chapter 4 called diffusion. Once there, it can touch the surface at many points at once. Each contact is weak, but together the contacts hold the protein in place, so a surface wetted by blood or tissue fluid does not stay bare.

The coat goes on changing after it forms. The two reviews mark three points along the way: the first proteins from blood within milliseconds, a layer on an implant within the first few minutes, and a stable layer found in living tissue after three hours. The marks describe different stages, and they come from reviews of different materials, blood contact in one and titanium implants in the other. They therefore give the order of events rather than one clock, and the first second of contact holds only the earliest of them.

Solved example: how much of the coat's forming falls in the first second
  • The start. The first proteins from blood adsorb "within milliseconds"; take one millisecond, a thousandth of a second, as the start.
  • The first second. One second is 1,000 milliseconds, three tenfold steps after the start.
  • The layer. A layer on an implant forms within the first few minutes. Two to five minutes are 120,000 to 300,000 milliseconds, about five tenfold steps after the start and about two after the first second.
  • The stable layer. Three hours are 10,800 seconds, or 10,800,000 milliseconds: about seven tenfold steps after the start and four after the first second.
  • The result. The first second spans three of the seven tenfold steps and holds the arrival of the first proteins. Measured from contact, the layer takes 120 to 300 times as long as the first second and the stable layer about 10,800 times as long, so most of the coat's forming happens after the first second.
  • Assumptions. One millisecond stands for "within milliseconds", and two to five minutes for "the first few minutes"; one minute would put the layer at nearly five tenfold steps. The marks come from reviews of blood contact and of implants, so the steps show the order and spread of the stages, not one measured clock.
22 — The coat a surface gathers, from milliseconds to hours
ORDER OF STAGES; THE THREE MARKED TIMES IN TENFOLD STEPS FROM 1 ms 1 ms 1 s 3 h a few minutes the first second BLOOD OR TISSUE FLUID MATERIAL on some surfaces, another protein takes the place of an early, abundant one water first; first proteins from blood within milliseconds a protein layer on an implant within the first few minutes on some surfaces, other proteins displace the early, abundant ones (the Vroman effect); no time given for the exchange adsorbed proteins unfold, more on water-repelling surfaces a stable layer found in living tissue after three hours; the cell meets the coat Schematic; shapes and counts illustrative. Marked times from two reviews of different materials (Kudzin and colleagues 2026, blood contact; Barberi and Spriano 2021, implants): the order of events, not one measured clock. On water-repelling surfaces that do not set off clotting, fibrinogen was not exchanged (Vroman and colleagues 1980).
Device materialTissue and cellsGas or fluid pathFocal detail
Plate 22 — Water reaches a surface first and proteins from blood begin to adsorb within milliseconds; a layer forms on an implant within minutes, and a stable layer was found after three hours. On some surfaces other proteins displace the abundant ones that arrived first, and adsorbed proteins unfold. The cell that arrives meets that coat, not the material.

The coat changes after it forms

On some surfaces the first proteins to arrive do not stay, because other proteins displace the abundant ones that adsorbed first. The review by Kudzin and colleagues calls this exchange the Vroman effect, and Barberi and Spriano add that proteins still in solution can take the places of those already held. Neither review gives a reason for the exchange. This book reads it as a contest of binding: a protein that takes another's place must hold that surface more firmly, so where the exchange runs, the coat shifts toward the proteins that bind best.

Earlier work by Vroman's group had found fibrinogen, a clotting protein, to be the main protein that plasma deposits on many man-made surfaces. In a 1980 bench study, Vroman and colleagues followed which plasma proteins occupied glass-like and other surfaces over time, using optical measurements of the coat and antibody tests that identify each adsorbed protein. On glass-like surfaces, which set off clotting on contact, the fibrinogen was rapidly replaced, mainly by high-molecular-weight kininogen and to a lesser degree by factor XII. On water-repelling surfaces that do not set off clotting, the fibrinogen was not exchanged.

The exchange came later in plasma lacking the kininogen, and slightly later in plasma lacking factor XII, which ties the exchange to those two proteins. The same plasma therefore leaves different coats on the two kinds of surface: mainly kininogen on a glass-like one, fibrinogen on a water-repelling one. Which protein holds the surface matters most in blood, because platelets, the cell fragments that start a clot, adhere preferentially to a coat of fibrinogen (Chapter 10).

Adsorbed proteins unfold

An adsorbed protein can lose its folded shape, and a protein that does so is said to denature. Its fold is held only by weak attractions, and a water-repelling surface offers the water-repelling groups tucked inside a place to go, so the protein can open out and lay them against the surface. Proteins on water-repelling, or hydrophobic, surfaces denature more than those on water-loving, or hydrophilic, ones, Barberi and Spriano report. The review by Kudzin and colleagues adds that fibrinogen can take extended forms on hydrophobic surfaces and on surfaces that form hydrogen bonds with it.

A protein that changes shape changes what it presents to cells. Because a water-repelling material tends to carry its proteins more opened out, the cells that arrive meet a different coat on it than on a water-loving material. Fibrinogen is one protein whose change of shape matters: once denatured on a surface, it is linked to the activation of platelets (Chapter 10).

Cells hold the coat, not the material

When a cell reaches the surface, it finds no bare material to touch. Cells interact with the adsorbed protein layer, not with the implant's surface, as Barberi and Spriano state. A cell holds the coat through integrins, receptor proteins that span its outer membrane. Fibronectin and vitronectin are two proteins that adsorb from body fluids, and both carry a run of three amino acids called RGD, after the one-letter codes of arginine, glycine and aspartic acid. Fibronectin's RGD site can bind one kind of integrin and activate it, and through such bonds the cell attaches.

The cell therefore reads the coat, and two materials can give it different things to read. Whether an RGD site faces outward depends on how its protein adsorbed and whether it unfolded. Whether fibronectin is there at all depends on the exchange that ran on that surface before the cell arrived. At the surface, the material acts on the cell through the proteins it adsorbs and the shapes they take, and whatever leaves the material acts as well (Chapter 4).

23 — A cell holds the proteins, not the material
CELL integrins: receptors across the cell's membrane MATERIAL RGD RGD fibronectin vitronectin unfolded: its sites changed or hidden no direct contact with the material the cell holds a protein, not the material Schematic. Cells meet the adsorbed layer through integrins; fibronectin can activate integrins through its RGD site, which vitronectin also carries (Barberi and Spriano, 2021 review).
Device materialTissue and cellsFocal detail
Plate 23 — A cell meets the adsorbed proteins, not the material. Its integrin receptors bind sites such as the RGD sequence of fibronectin and vitronectin, so what the cell holds depends on which proteins adsorbed and how they unfolded.

What the surface decides

Wettability is how readily water spreads on a surface. A drop of water spreads flat on a hydrophilic surface and beads up on a hydrophobic one, and the angle at the drop's edge measures the difference. Wettability shapes how much protein a surface takes up as well as how far that protein unfolds: hydrophobic surfaces adsorb more protein than hydrophilic ones, in Barberi and Spriano's review. Implantable materials are predominantly hydrophobic, Sheikh and colleagues note in a 2015 review of the body's response to implants. The same review adds that changing a surface's chemistry, wettability or form changes both the proteins it adsorbs and the response of cells.

A surface's charge changes how readily its proteins are displaced. Albumin, a blood protein, is displaced faster on negatively charged titanium than on positively charged alumina at pH 7, which is close to the body's own. Roughness mattered less: over the range of the studies that Barberi and Spriano review, roughness changed only slightly how much albumin from cattle a surface adsorbed. Topography is a surface's shape at the scale of cells, and Sheikh and colleagues count it among the properties that set the body's response. Chemistry can also be set on purpose, because attaching chosen chemical groups to a surface can change what it adsorbs.

These findings are examples, not laws: studies of adsorption often use surfaces that "differ only in one or two properties", as Sheikh and colleagues note. By contrast, one molded part differs from another in chemistry, wettability, charge and form at once, and its coat follows from all four together and from the fluid that wets it. For that reason the effects of a coat are read by tests on the part itself, in blood or in tissue (Chapters 9 and 10), and not from one property on a datasheet.

A surface acts without releasing anything

The coat completes the invariant this book rests on, that no material is inert at the scale the body works on. Chapter 4 showed how molecules leave a material, and this chapter shows that proteins go onto the surfaces that body fluids wet. A material that released nothing at all would still be coated, the proteins on it could still be exchanged or unfolded, and cells would still respond to what it held. A response can therefore begin with no molecule leaving the material. An extract collects what leaves and holds nothing of the coat, so the chemistry of Part V needs the tests of Chapter 18 beside it.

The running device shows where the coat matters. Its mask cushion is silicone and rests on intact skin (Chapter 2). The skin's outer layer keeps blood and tissue fluid away from the cushion, so the plasma proteins of this chapter do not reach the silicone. The cushion's questions are therefore what leaves it and what the skin takes up, which Chapter 11 answers. A tube of the same silicone carrying blood would gather the whole coat, because the coat depends on the fluid that wets a surface as well as on the material. The device's foam sits in the air path and touches no body fluid in place, but particles shed from it can reach the airway, and Chapter 9 follows them there.

In practice

List each part of a device with the body fluid that wets it in use: blood, tissue fluid, or neither, as for a part resting on intact skin. For each wetted part, describe the surface as it leaves the last process step, not only the resin: its wettability, its charge where known, its topography and any coating or surface treatment. Treat a change to any of these as a change to what the body meets, even when the resin stays the same (Chapter 23). For the running device, record in the evaluation that no part in place is wetted by blood or tissue fluid, and why (Chapter 22), and assess the particles shed into its air path separately (Chapters 9 and 19).

+ What this chapter established
  • Water reaches a surface first; reviews of different materials put the first proteins from blood within milliseconds, a layer on an implant within minutes and a stable layer in living tissue after three hours.
  • The coat changes after it forms: on some surfaces other proteins displace the abundant ones that arrived first (the Vroman effect), and adsorbed proteins unfold, more on hydrophobic surfaces.
  • Cells meet the adsorbed proteins, not the material, and hold them through integrin receptors that bind sites such as the RGD sequence of fibronectin.
  • Wettability, charge, topography and chemistry shape the coat, so a material that releases nothing still draws a response wherever blood or tissue fluid wets it, while a cushion on intact skin meets no such fluid.