Contact, counted.
+ The questionHow do the kind and the length of contact decide what must be shown?
Four groups of tissue
Three years after starting treatment, the median user in a Norwegian study still wore a breathing mask for about 6 hours a night. In that 2022 prospective study, Møkleby and Øverland followed 163 patients who began the treatment at one sleep center, read the hours of use from the machines' own records, and still had data for 99 of them at long-term follow-up. Users who stopped or were lost were not counted, so the figure describes those who kept the machine. Many do not keep it. In a 2021 analysis of French national insurance claims, Pépin and colleagues found that 47.7 % of 480,000 new users had stopped within three years. ResMed, a device maker, funded the analysis, and three of its employees were among the authors. For those who continue, the mask touches the face for a quarter of every day, for years.
How that contact is judged depends first on the tissue it meets. ISO 10993-1, the standard for the biological evaluation of devices, sorts each component of a device by what it touches. Its 2025 edition has four groups, as testing laboratories report it: intact skin; intact mucosal membranes, the moist linings of passages such as the nose and mouth; breached or compromised surfaces and internal tissues other than blood; and blood. Bone and the soft tissue around an implant are internal tissues, so an implant falls in the third group, or in the fourth where it meets blood. The groups follow the tissue that each component touches rather than where the device sits. That is a change from earlier editions, and Chapter 21 traces it.
In the terms of Chapter 1's chain, the four groups can be read as routes that differ in what stands between a device and living tissue. On intact skin, whatever leaves a material must cross the skin before it reaches living cells. A mucosal lining is a barrier too, but a moist and living one, so what reaches it can pass into the tissue beneath. Where a surface is breached, or a device sits inside the body, a release meets living tissue with nothing in between. In blood, whatever enters travels through the whole body, as Chapter 1 showed for a blood bag. The word "intact" in the first two names states the barrier's condition: the same cushion pressed on broken skin would belong to the third group.
The breathed air has a standard of its own
The air that a breathing device delivers fits none of the four groups. The parts that carry it need not touch the patient at all, yet whatever they release travels with the air onto the airway and into the lungs. A separate series of standards, ISO 18562, covers this gas pathway. Its first part sets out the evaluation within a risk management process. The next three test the delivered air for particles and for volatile organic compounds, and the water that condenses in the circuit for the substances it carries, as a test laboratory describes the series. Chapter 19 sets out the limits that these tests are judged against.
The series' first edition appeared in 2017, and a record from the year before shows how a gas pathway was sorted without it. In a 2016 summary to FDA for a hospital humidification chamber, its maker put the chamber's contact under ISO 10993 as tissue reached "by way of gas path", with direct contact with mucosa, and listed six kinds of test for 14 days of use. ISO 18562-1 also counts duration, in bands of its own. Its 2017 edition named them limited, up to 24 hours; prolonged, up to 30 days; and permanent, 30 days or more, in the accounts of a test laboratory and of an instrument maker that sells the equipment for the tests. The series' 2024 edition adds the terms "infrequent use" and "total exposure period", in another laboratory's list of its changes.
How the nights add up
Duration is the second axis, and ISO 10993-1 divides it into three bands. Its 2025 edition keeps limited contact, up to 24 hours; prolonged contact, from 24 hours to 30 days; and long-term contact, more than 30 days, in a laboratory's account of the edition. FDA's 2023 guidance uses the same three. A single use is easy to place, but a mask worn for 6 hours a night is not, because each night is short and the nights repeat.
Repeated contact is counted because its effects can add up. FDA's guidance describes exposure as "a one-time exposure, a constant exposure over time, or an intermittent exposure over time that could have a cumulative effect". Slides from FDA's town hall of September 2026 name bioaccumulation, a substance building up in the body, among the bases for setting duration. In the 2025 edition, repeated contact is added up in days rather than hours, as testing laboratories report it. The edition's own foreword states that content "has been added to provide guidance and clarification of calculation of exposure duration", and the same FDA slides list intermittent contact among the edition's new terms.
Testing laboratories reproduce the edition's definitions of these terms in their notes on it. In the note by NAMSA, a contract research organization, a contact day is a "Day in which a medical device or component thereof comes into contact with tissues, including circulating blood, irrespective of the length of time of that contact within the day". The same note gives the total exposure period as the "Number of contact days between the first and last use". Under the two definitions, a few minutes of contact count as a whole day, and two days of use already make prolonged contact. The note puts the change plainly: "Gone are the days of adding together the seconds of multiple exposure".
The laboratories agree that a part-day counts as a day, and they differ in the details. Complife counts the calendar days from first to last use for repeated or continuous use, and only the days of actual contact for intermittent use, in which at least 24 hours pass between contacts. NAMSA counts contact days between the first and last use. Labcorp set the threshold for a contact day at contact of more than 1 minute, "(exceptions apply)", where NAMSA's definition holds for contact of any length; Labcorp wrote its note before the edition was published. Complife also names the running device's kind of use: it places "CPAP devices used for a few hours per day" among those that "may fall within prolonged or long-term contact categories".
- The pattern. A user wears the mask for 6 hours every night, the median at long-term follow-up in the Norwegian study.
- Hours added up. Long-term contact begins after 30 days, or 30 times 24, which is 720 hours. At 6 hours a night, 720 divided by 6 is 120 nights, so the 121st night passes into long-term contact.
- Contact days. Each night is one contact day, whatever its length, so the 31st night passes 30 days.
- The result. Counted in contact days, the cushion is in long-term contact after about a month of nightly use rather than about four months. 121 divided by 31 is about 4, so contact days reach the boundary about 4 times sooner.
- Fewer nights. Medicare's minimum use for coverage, 4 hours or more on 70 % of nights, gives 21 contact days in 30. The 31st contact day then comes on about the 44th calendar day, 31 divided by 0.7. On Complife's calendar reading for repeated use, the boundary is the 31st calendar day. Either way the cushion is in long-term contact within about six weeks.
- Assumptions. The counting rules are the laboratories' readings of the standard's clause. The 6-hour median describes users who kept the machine, and the study did not report the share of nights used. The 70 % figure is a minimum for coverage, not observed use.
What each group and band asks for
The tissue group and the duration band together decide which endpoints an evaluation must address. An endpoint is one biological effect, such as harm to cells, allergy or damage to genes. The 2025 edition calls them biological effects and sets them out in four tables, one for each type of contact, where the 2018 edition had one matrix, as three laboratories report (Chapter 21). Within a table, a longer contact adds effects. Measurlabs, a testing laboratory, reports that a genotoxicity evaluation is "now required" for every device with prolonged or long-term contact, and that carcinogenicity now extends to long-term contact with intact mucosa. NAMSA, by contrast, calls genotoxicity a "required endpoint of consideration".
The tables list what an evaluation must consider, not tests to run. In a 1997 trade-press review of the standard as it then stood, Bollen and Svendsen wrote that its matrix "should not be considered as a checklist for the different tests that have to be performed". The 2025 edition places the choice of tests inside a biological risk estimate (Chapters 18 and 21). FDA's town hall slides name the step that this chapter has followed: "Define contact category, tissue type, duration (limited, prolonged, longterm), and frequency". In the 510(k) summary of one recent breathing machine, its maker gives the console "permanent contact duration (>30 days) including dry and humidified gas pathways", and lists tests under three parts of ISO 18562 and for harm to cells, allergy and irritation.
The running device, sorted
Sorted this way, the running device of Chapter 1 has three kinds of contact: the air it delivers, the condensate in its circuit and intact skin. Its gas pathway, from the intake to the mask, is evaluated under ISO 18562, and for a patient who keeps the machine for years its contact is long-term on any count. The water chamber belongs to the same pathway. Chapter 3 makes it of polycarbonate, and what heated water draws from that plastic can reach the patient in the condensate that the series' fourth part tests. The silicone cushion and the fabric headgear touch intact skin every night. One maker states a useful life of 6 months for its nasal mask, about 180 nights, which is long-term contact whether the nights are counted as days or their hours are added up.
The housing, the display and the outside of the hose touch the user's hands for moments at a time. Whether such a touch makes a contact day is one of the details the laboratories' notes leave open, since NAMSA counts contact of any length and Labcorp set a threshold of more than a minute. This book treats these parts as brief contact with intact skin, and the chapters that follow evaluate the gas pathway, the water chamber and the cushion.
Sort every part of a device by the tissue it touches and by its duration before choosing any test. Count repeated use in contact days, state in the evaluation which reading of the counting rule it follows, and take the longest foreseeable use, such as a mask's stated useful life. Put every part in the path of the breathed air under ISO 18562 as well, whatever else it touches.
- ISO 10993-1:2025 sorts contact into four tissue groups: intact skin; intact mucosa; breached surfaces and internal tissues; and blood, while the breathed air of a gas pathway is evaluated under ISO 18562.
- Contact is limited up to 24 hours, prolonged up to 30 days and long-term beyond, and laboratories read the 2025 edition as counting a part-day as a whole contact day, though they differ on the shortest contact that counts.
- Counted in contact days, a mask worn for 6 hours a night passes into long-term contact on the 31st night, about four times sooner than if its hours were added up.
- The tissue group and the duration band decide which biological effects an evaluation must consider, and a longer contact adds effects such as genotoxicity.