What a material holds, and what leaves it.
A material in a device is a mixture: a polymer, the additives that make it mold and last, and the leftovers of its making. The additives and leftovers can move through the polymer and out of it. Chapter 3 takes the mixture apart, family by family, and asks what a supplier's certificate says about it. Chapter 4 follows the molecules that leave a solid into blood, saline or air, and shows why the liquid and the contact time decide how much arrives. Chapter 5 turns to the polymer itself, which water, heat, oxygen and ozone can cut into new, smaller molecules, and explains why two polyurethane rubbers that look alike can last a few weeks or a year in hot water.
A plastic is a recipe.
+ The questionWhat is in a medical plastic besides the polymer?
Chains, and what lies between them
A medical plastic is a polymer together with the substances added to make it mold, last and look right, and with what its making leaves behind. The polymer is the frame: a long chain built by joining small molecules, the monomers, end to end many times over, and a molded part is a mass of such chains lying tangled together. Almost everything else in the part lies among the chains rather than in them. The whole mixture is the plastic's recipe. The recipe is part of the material as made, the first of the five links that Chapter 1 traces from material to response, and it fixes what a part can release.
Every batch of a polymer holds chains of many lengths, down to chains of only a few units, called oligomers. Short chains are small enough to move through a plastic as additive molecules do, and they are more than a trace. In PET, the polyester of drink bottles, ring-shaped oligomers make up 0.06 to 1 % of the material, depending on the type, as much as many stabilizers (next section). Hahladakis and colleagues give that figure in a 2018 review of the additives in plastics, and they note that monomers and oligomers both tend to migrate from packaging into food. The review was written mostly about packaging, and its authors call it neither systematic nor comprehensive. A chain can also break in use, and each break makes a new molecule (Chapter 5).
Additives sit between the chains, each with a job
An additive is a substance mixed into the polymer to give the plastic a property the chains alone lack. In nearly all cases it is "not chemically bound to the plastic polymer", in the words of the 2018 review. The exception is a reactive additive, which joins the chain as the chain forms. An unbound additive is held only by its attraction to the chains around it, so it can leave once the part touches something that dissolves it (Chapter 4).
Each class of additive answers a weakness of the bare polymer, and the 2018 review tabulates the typical content of each. PVC is rigid alone, so it takes plasticizers, at 10 to 70 % of a plastic's weight, which improve "the flexibility, durability and stretchability of polymeric films". Chains degrade by oxidation, including oxidation that ultraviolet light starts, and antioxidants and ultraviolet stabilizers, at 0.05 to 3 %, delay it. PVC also degrades at high temperature, and heat stabilizers prevent that. They are used mainly in PVC, at 0.5 to 3 %. A surface can stick in its mold, and slip agents, at 0.1 to 3 %, lower its friction and help the part come free. An additive's share runs from a trace to most of the part: plasticizers can reach 70 %, fillers 50 %, and organic pigments can be as little as 0.001 %. The share does not decide how readily an additive leaves: soluble colorants migrate easily, and organic pigments little.
The making leaves residues
Monomers, intermediates and catalysts used in making a plastic form a separate group: the 2018 review does not count them as additives. The European Commission's scientific committee on emerging health risks, SCENIHR, assessed bisphenol A, or BPA, in medical devices in a 2015 opinion. Unreacted monomers "are almost always released from polymer resins", the opinion states, and devices made of polycarbonate or polysulfone "may contain BPA residues", since BPA is a building block of both. Time, high temperature and high pH raise the release. A residue limit can be very small. For PVC, the 2018 review gives the EU's limit for residual vinyl chloride, its monomer, in materials that touch food: 1 mg per kg of plastic.
Catalysts, which speed the reactions that join the units without being used up, can stay in the part as well. Polyurethane foams for medical use are made with tin or tertiary-amine catalysts, and some of these amines are toxic, as Grzęda and colleagues note in a 2023 bench study. The study compared amine catalysts for a soft foam meant for biomedical use and measured the foams' thermomechanical properties and the toxicity of their extracts to cells. One of the six authors worked for a company, Fampur Adam Przekurat. Reactive amines bind into the polymer, as the reactive additives of the previous section do, so they cannot diffuse out of the foam, the authors note. The study did not measure how much catalyst remained in its foams.
- The plasticizer. 1 % by weight is 10,000 mg per kg, so at the review's typical 10 to 70 % a flexible PVC holds 100,000 to 700,000 mg of plasticizer per kg.
- The antioxidant. At the 0.05 to 0.10 % that two staff of an additive maker give for most polyolefins, a polyolefin holds 500 to 1,000 mg per kg.
- The residue. The EU's food-contact limit for residual vinyl chloride in PVC, as the 2018 review gives it, is 1 mg per kg, one part in a million.
- The result. In PVC the plasticizer content is 100,000 to 700,000 times the residue limit, five to nearly six tenfold steps. A polyolefin's antioxidant, set beside the same limit, is 500 to 1,000 times it, about three steps. A laboratory may need to see one part in a million beside an ingredient that makes up a tenth or more of the weight.
- Assumptions. The ranges are typical for plastics in general, mostly packaging; the antioxidant range is an additive maker's; and 1 mg per kg is a food-contact limit, not a device limit. A share does not decide whether an ingredient matters: the dose and the substance do (Chapters 14 and 15).
Each family carries its own recipe
PVC takes about 80 % of all plasticizers used, in the 2018 review's estimate, and they make it the soft plastic of bags and tubing. DEHP leads the plasticizers by volume, and Chapter 4 follows it into stored blood. In 2004, companies supplying PVC and plasticizers for devices told a trade magazine that calcium-zinc heat stabilizers were typically used in medical-grade PVC. Barium-zinc, which they described as better at protecting against heat, was kept out of devices because of its heavy metals.
Polyethylene and polypropylene, the polyolefins, carry antioxidants instead. Most polyolefins contain one or more antioxidants at 0.05 to 0.10 %: a hindered phenol as the primary antioxidant, with a phosphite as the secondary one. Van Beusichem and Ruberto give these levels in an undated poster. Both worked for Ciba, which then made the antioxidants the poster names. Ruberto's 2010 trade-journal review of the plastics in single-use bioprocess equipment adds stabilizers against heat and light, lubricants and mold-release agents, and it mentions residual catalyst without naming the catalyst. Additives often arrive as a masterbatch, a concentrate at 5 to 30 % or more, so their share in a part depends on how far the concentrate is diluted into the plain polymer. Polycarbonate may carry its monomer, BPA, as a residue (previous section).
A polyurethane forms when an isocyanate reacts with a polyol, a molecule carrying several alcohol groups, and each such reaction makes a urethane link, the bond Chapter 5 returns to. A foam also needs a blowing agent, which makes the gas that swells the setting mixture into cells. The foams in the study by Grzęda and colleagues used water as the blowing agent, with a silicone surfactant, a soap-like ingredient that steadies the cells as they form.
Silicone elastomers, the rubber of the running device's mask cushion, are built from chains of silicon and oxygen carrying methyl groups and are reinforced with fine silica. Curing turns a liquid silicone into a rubber by joining its chains to one another into a network. ASTM's guide to silicones for medical use names three ways to cure them: a platinum catalyst, an organic peroxide or a tin compound. ASTM issued the guide in 2000, reapproved it in 2011 and has since superseded or withdrawn that version. One peroxide, 2,4-dichlorobenzoyl peroxide, leaves small amounts of polychlorinated biphenyls, compounds that carry chlorine, and other by-products, and a further bake, the post-cure, removes them. Silicone parts also hold short chains, in amounts that depend on process and storage, and stripping or post-cure reduces them.
"Medical grade" is a supplier's word
A resin sold for devices may be called medical grade, and the term records choices its supplier made, not a clearance or approval by FDA. FDA "does not clear or approve individual materials that are used in the fabrication of medical devices", its 2023 guidance on ISO 10993-1 states. The guidance adds that the risk assessment "should evaluate the final finished device", including the methods of manufacture and any residues of manufacturing aids. A medical grade is therefore a recipe with documents attached, and a certificate of biological tests on the resin is one of the documents.
One such certificate is USP Class VI, which industry adopted "as the default", in the words of a 2024 technical article by the supplier MilliporeSigma. A Class VI certificate reports three tests in animals that chapter 88 of the United States Pharmacopeia set out. An undated trade article describes six classes, from Class I, the simplest, to Class VI, the most rigorous. As that article and MilliporeSigma's describe the tests, the material is extracted in saline, alcohol in saline, polyethylene glycol and sesame oil. The extracts are injected into mice, which are watched for three days for changes in body weight or death. Extracts are also injected into the skin of rabbits, and the sites are scored over 72 hours for reddening, scabbing and swelling. Rods of the material are implanted in rabbit muscle for seven days beside control rods of high-density polyethylene, and the capsule that forms around them is measured.
A certificate's tests run for three to seven days, while the running device's user wears a mask every night for years (Chapter 2). They also test the resin before molding, bonding, cleaning and aging, and FDA's guidance states why that matters: "manufacturing and processing could affect the final chemistry presented to the body". Chapter 6 follows what processing adds. The evaluation of the finished device follows ISO 10993-1, whose sixth edition was published in November 2025, although FDA's 2023 guidance addresses the 2018 edition (Chapter 21).
MilliporeSigma's 2024 article calls Class VI "the most stringent medical device standard", but chapter 88 belongs to a pharmacopeia. In November 2026 USP's page for the chapter, which cites the 2024 United States Pharmacopeia, describes only a systemic injection test, for the packaging of drugs, and refers devices to FDA's device guidance. USP has also revised chapter 88 to remove Classes I to VI from 1 December 2026, a trade column of October 2026 reports. A Class VI result is one input to the evaluation of the finished part, as are the cell tests of the companion chapter 87, or of ISO 10993-5, when they are run on a resin (Chapter 18).
The running device, listed by family
The nasal mask of Chapter 1's running device takes its materials from a mask that FDA cleared in 2020. The maker of that mask names them in its 510(k) summary, its own account to FDA. The frame, elbow and swivel are polycarbonate, and the cushion is silicone on the side that touches the face, mounted on polycarbonate. The headgear is polyurethane foam, nylon and polyester. The sound-damping foam beside the blower is a polyurethane foam, and Chapter 5 chooses its type, polyester or polyether, from how each breaks down in a warm, humid air path. Metals and ceramics fall outside this chapter, because the running device's parts that this book follows are all polymers.
Several makers' public summaries for such devices name no material for the water chamber or for the cuff that joins the hose to the machine, so this book chooses both by family. The chamber is polycarbonate, stiff and clear so that the water level shows, and it holds water heated night after night. The hose and its cuff are a thermoplastic elastomer, a soft, rubber-like plastic that melts for molding. A bill of materials that names only families leaves out what each part brings: the cushion its short chains and cure residues, the foam what remains of its catalysts and surfactant, and the chamber any BPA residue its polycarbonate may hold.
A resin maker can keep its formulation from the device maker and still let FDA review it, in a master file (Chapter 23). The information in a master file alone "may not be sufficient to establish the biocompatibility of the device", FDA's guidance states. The guidance adds that a change of resin supplier can also affect biocompatibility, since a new supplier may not remove processing solvents. The device maker may know the recipe only as far as the supplier's documents state it.
For each part that touches the user or the air path, record the polymer family, the grade, and what the supplier states about additives, residues and processing aids. Read each certificate for which tests it covers, on which material, extracted in what, and when. Plan the evaluation on the finished part as molded, cleaned and aged (Chapters 6 and 7), with the resin's results as inputs. Agree with each supplier how a change of recipe will be notified (Chapter 23).
- A medical plastic is a polymer with short chains of its own, additives that in nearly all cases are not bound to it, and residues of its making such as monomer and catalyst.
- Each additive answers a weakness of the bare polymer, and in PVC the plasticizer content stands five to nearly six tenfold steps above the food-contact limit for residual monomer.
- Each family brings its own recipe, from plasticizer and heat stabilizer in PVC and antioxidants in polyolefins to cure residues and short chains in silicone.
- "Medical grade" and a USP Class VI certificate describe a resin as its supplier tested it, and FDA expects the evaluation of the finished device as made.